Measurement of the Number of Photoelectrons per MeV for Neutron Detectors

dc.contributor.authorİren, Emre
dc.contributor.authorÖzok, Ferhat
dc.contributor.authorErduran, Mustafa Nizamettin
dc.contributor.authorErduran, Mustafa Nizamettin
dc.date.accessioned2024-01-05T11:52:00Z
dc.date.available2024-01-05T11:52:00Z
dc.date.issued2021
dc.departmentMühendislik ve Doğa Bilimleri Fakültesi
dc.description.abstractThe light gains of a detector system (photomultiplier tube and scintillator) is characterized by the number of photoelectrons produced per unit of energy (usually MeV). In this study, the results of light gain tests for 10 Hamamatsu R11833 model photomultiplier tubes are shown. Various radioactive sources were used in conjunction with the EJ309 liquid scintillator to determine the light gain. The number of photoelectrons produced per unit energy was determined by comparing the size of the signal generated by the photons produced by photomultiplier tube during the scintillation process using radioactive sources and the size of the signal generated when a single photoelectron was produced by the photomultiplier tube. The test results show that the performances of the photomultiplier tubes are sufficient for neutron detector arrays used in experiments with radioactive ion beams. The supply voltage of the photomultiplier tubes must be changed to achieve similar light gains.
dc.identifier.citationİREN, E., ÖZOK, F., & ERDURAN, M. N. (2021). Measurement of the Number of Photoelectrons per MeV for Neutron Detectors. European Journal of Technique (EJT), 11(2), 136-139. https://doi.org/10.36222/ejt.950944
dc.identifier.doi10.36222/ejt.950944
dc.identifier.endpage139
dc.identifier.issn2536-5010
dc.identifier.issn2536-5134
dc.identifier.issue2
dc.identifier.orcidEmre İren |0000-0002-5751-7479
dc.identifier.orcidFerhat Özok |0000-0001-9630-7362
dc.identifier.orcidMustafa Nizamettin Erduran |0000-0003-0852-9753
dc.identifier.startpage136
dc.identifier.trdizinid1123573
dc.identifier.urihttps://doi.org/10.36222/ejt.950944
dc.identifier.urihttps://hdl.handle.net/20.500.12436/5580
dc.identifier.urihttps://search.trdizin.gov.tr/tr/yayin/detay/1123573
dc.identifier.urihttps://dergipark.org.tr/en/download/article-file/1818431
dc.identifier.volume11
dc.indekslendigikaynakTR-Dizin
dc.institutionauthorErduran, Mustafa Nizamettin
dc.language.isoen
dc.publisherHibetullah Kılıç
dc.relation.ispartofEuropean Journal of Technique (EJT)
dc.relation.publicationcategoryMakale - Ulusal Hakemli Dergi - Kurum Öğretim Elemanı
dc.rightsinfo:eu-repo/semantics/openAccess
dc.subjectPhotomultiplier Tube
dc.subjectSingle Photoelectron
dc.subjectLiquid Scintillator
dc.subjectPhotoelectron Yield
dc.subjectNeutron Detection
dc.titleMeasurement of the Number of Photoelectrons per MeV for Neutron Detectors
dc.typeArticle
dspace.entity.typePublication
relation.isAuthorOfPublicationf15c7b3b-1513-45ba-ba89-0994e82a74cb
relation.isAuthorOfPublication.latestForDiscoveryf15c7b3b-1513-45ba-ba89-0994e82a74cb

Dosyalar

Orijinal paket

Listeleniyor 1 - 1 / 1
Yükleniyor...
Küçük Resim
İsim:
10.36222-ejt.950944-1818431.pdf
Boyut:
1.42 MB
Biçim:
Adobe Portable Document Format
Açıklama:
Makale dosyası / Article file

Lisans paketi

Listeleniyor 1 - 1 / 1
Yükleniyor...
Küçük Resim
İsim:
license.txt
Boyut:
1.44 KB
Biçim:
Item-specific license agreed upon to submission
Açıklama: